Home  >  Electronic Measuring Instruments  >  Industrial Metal Detectors

coating thickness gauge - TT260

1.magnetic induction (F) and eddy current (N), 2.6 types of probes are available for various applications, 3.continuous/single - more info check here: http://www.buymeasuringtools.com/coating-thickness-gauge-10048451/

Quick Brief

Place of Origin: Liaoning China (Mainland) Brand Name: TIMES Model Number: TT260 color: dark gray

Product Features:

1.magnetic induction (F) and eddy current (N)

2.6 types of probes are available for various applications

3.continuous/single

Business Terms:

Port: Tianjin

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 10 Set/Sets per Day

Payment Terms: T/T

Contact Supplier
coating thickness gauge
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: carton/plywood

Delivery Detail: 3-5 days

Specifications and Product Details:

Two measuring methods of paint coating thickness gauge: magnetic induction (F) and eddy current (N)

Two switch off modes: manual and auto

Two switch off modes: manual and auto

Technical Specification
Measuring range

Refer to the table below

Probes available
Tolerance
Minimum resolution:
Measuring condition
Operation languageEnglish
StandardsDIN, ISO, ASTM,BS
CalibrationZero and foil calibration
StatisticsNumber of measurements, mean, standard deviation,
maximum and minimum of 3000 readings
Data memory495 readings
LimitsAdjustable with alarm
InterfaceRS-232
Working temperature0-40°C
power supplyNicd rechargeable batteries 1.25V
Dimensions270mm×86mm×47mm
WeightApprox. 530g
Optional probes and technical specification

Probe model

F400

F1F1/90ºF10N1

CN02

Operating principle

Magnetic induction

Eddy current

Measuring range (µm)

0-400

0-1250

0-100000 to 1250 μm
0 to 40μm (for chromeplate on copper)

10~200

Low range resolution (µm)

0.1

0.1

100.1

1

Accuracy

One-point calibration (µm)

±(3%H+1)

±(3%H+10)±(3%H+1.5)

±(3%H+1)

Two-point calibration (µm)

±[(1~3)H%+0.7]

±[(1~3)H%+1]

±[(1~3)%H+10]±[(1~3)%H+1.5]

Measuring conditions

Min curvature of the min area (mm)Convex

1

1.5Flatten103

Flatten

Diameter of the min area (mm)

φ3

φ7φ7φ40φ5

φ7

Critical thickness of substrate (mm)

0.2

0.50.520.3

unlimited

Standard delivery

  • Main unit
  • Calibration foil set
  • Substrate
  • Charger
  • Probe
  • Instruction manual
  • TIME certificate
  • Warranty card