Quick Brief

Place of Origin: China (Mainland) Brand Name: Jiechuang Model Number: JC-ZJF-3 display: blue screen

Product Features:

1.Blue screen display, can be used all-weather.

2.Has a ferquency(Hz) measurement and emulation function.

Business Terms:

Port: Shang Hai

Minimum Order Quantity: 1 Unit/Units

Supply Ability: 1000 Unit/Units per Week

Payment Terms: L/C,T/T

Contact Supplier
calibration instrument
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: carton and plastic bag

Delivery Detail: 15~30 days

Specifications and Product Details:

JC-ZJF-3 process signal simulation calibration instrument is specially design for industry process control, intelligence menu prompt, digitization operation, kompleta function, 18 input, output of thermal measurement, synchronous operation, independent operation, and isolate from each other. Possess this instrument, you can easy to complete the routine maintenance and calibration.

Characteristic

  1. Blue screen display, can be used all-weather.
  2. With the signal output and measurement of V, mV, mA, Ω and thermocouple (TC), resistance temperature (RTD).
  3. Has a ferquency(Hz) measurement and emulation function.
  4. 24V DC power distribution provide the loop power for transmitter, and can measure the output value of mA form transmitter in a same time.
  5. Measure the thermocouple (resistance temperature), display rhe mV(Ω), degree celsius in a same time, not need to check the reference table any more.
  6. Using the high-precision temperature sensor to provide for cold junction compensation for thermocouple.
  7. All the signals can trimming or continuous add or subtract in arbitrary decimals.
  8. Has the function of simulation transmitter, can calibration DCS channel 4-20 mA loop current.
  9. Has RS232 comminication interface, can realize the fast revise of instrument out-of-tolerance though the PC.

Output index

OutputRangeResolutionPrecision(%reading+%full range)LoadNotes
Voltage(mV)0-±99.9991μv0.02%+0.015%200Ω(min)DC
Voltage(V)0-±9.99990.1mv0.02%+0.02%200Ω(min)DC
Current(mA)0-±25.0001μA0.02%+0.025%20Ωless than R less than 50ΩDC
Artificial resistance(Ω)0-999.990.01Ω0.02%+0.01%R×1less than10V

50μAless than 1less than 10 mA

Voltage(V)24±10%30mA(max)Instrument power
Frequency(Hz)0-1 k0.01Hz0.02%+0.02%10KΩTTL

resistance temperature measure and output accuracy

The type of thermocouple

Range of measurement

The range of give accuracy(degree celsius)

Accuracy

MeasurementOutput

Pt100

-50~600

-50~2000.20.2
200~6000.250.2
Cu100-50~150-50~1500.50.5
Cu50-50~150-50~1500.50.5

Measure index

MeasureRangeResolutionAccuracy(%degree+% full range)Input impedanceNotes
Voltage(mV)0~±100.0001μV0.02%+ 0.02%10-9ΩDC
Voltage(V)0~±10.00000.1mV0.02%+0.25%1MΩDC
Voltage(V)0~±50.000mV0.02%+0.025%1MΩDC
Current(mA)0~±25.00000.1μV0.02%+0.025%50ΩDC
Resistance(Ω)0~999.990.01Ω0.02%+0.02%
Resistance(Ω)0~9999.90.1Ω0.02%+0.02%
Frequency(Ω)0~10K0.1Hz0.02%+0.02%10KΩTTL