Home  >  Optical Instruments  >  Spectrometers

UV/VIS Spectrophotometer - UV-2601

UV/VIS Spectrophotometer(UV-2601), 1. Double beam, 2. 190-1100nm, 3. 2nm(5nm, 4nm, 1nm optional), 4. PC control, USB port, LCD - more info check here: http://www.buymeasuringtools.com/UV-VIS-Spectrophotometer-10268536/

Quick Brief

Place of Origin: Beijing China (Mainland) Brand Name: REYLEIGH Model Number: UV-2601 Size: 630*470*210mm Weight: 26kg Wavelength Range: 190-1100

Product Features:

UV/VIS Spectrophotometer(UV-2601)

1. Double beam

2. 190-1100nm

3. 2nm(5nm, 4nm, 1nm optional)

4. PC control, USB port, LCD

Business Terms:

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 50 Set/Sets per Month

Payment Terms: L/C,T/T

Contact Supplier
UV/VIS Spectrophotometer
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Carton

Delivery Detail: Two Weeks

Specifications and Product Details:

FEATURES:

UV-2601 double beam UV/VIS spectrophotometer successfully realizes high accuracy and reliability measurement to meet various application requirements. It can be used extensively for qulitative and quantitative analysis in such fields as biochemical research and industry, pharmaceutical analysis and production, education, environmental protection, food industry, clinical examination, sanitation and antiepidemic etc..

1 Wide wavelength range, satisfying requirements of various fields

2 Five options for spectral bandwidth selection, 5nm, 4nm, 2nm and 1nm, made according to customer's need and satisfying the requirements of pharmacopoeia

3 Fully automated design, realizing the simplest measurement

4 Optimized optics and large scale intergrated circuits design, light source and receiver from world famous manufacturer all add up to high performance and reliability

5 Rich measurement methods, wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirements.

6 Automatic 10mm 8-cell holder, changeable to automatic 5mm-50mm 4-position cell holder for more choices.

7 Data output can be obtained via a printer port and a 8 Parameters and data can be saved in case of power failure for user's convenience.

8 PC controlled measurement can be achieved via RS-232 interface (USB port) for more accurate and flexible requirement.

SPECIFICATIONS:

Wavelength Range:190-1100nm
Spectral Bandwidth:2nm (5nm, 4nm, 1nm optional)
Wavelength Accuracy:±0.3nm
Wavelength Reproducibility:≤0.15nm
Optical System:Double beam, Auto scan, dual detectors
Photometric Accuracy:±0.3%T(0~100%T), ±0.002A(0~1A)
Photometric Reproducibility:≤0.15%T
Working Mode:T, A, C, E
Photometric Reproducibility:-0.3~3.5A
Stray Light:≤0.1%T(NaI, 220nm, NaNO2 340nm)
Baseline Flatness:±0.002A
Stability:≤0.001A/h( at 500nm, after warming up)
Noise:±0.001A( at 500nm, after warming up)
Display:6 inches high light blue LCD
Detector:Silicon photo-diode
Power:AC 220V/50Hz, 110V/60Hz, 180W
Dimensions:630×470× 210 mm
Weight:26kg

UV/VIS Spectrophotometer suppliers    UV/VIS Spectrophotometer manufacturers