Home  >  Pressure Measuring Instruments  >  Other Pressure Measuring Instruments

Temperature Calibrator HS211 - HS211

Temperature Calibrator HS211 intergrates Volt, current, RTD, frequency,thermocouple, pressure measurement and output into one - more info check here: http://www.buymeasuringtools.com/Temperature-Calibrator-HS211-10309030/

Quick Brief

Place of Origin: Beijing China (Mainland) Brand Name: Huaxin Model Number: HS211 Using environment temperature: (-5-45)℃ Relative humidity: 20% - 90% No condensation Store environment temperature: (-20-55)℃ Dimension: 355x285x140mm Weight: 5KG

Product Features:

Temperature Calibrator HS211 intergrates Volt, current, RTD, frequency,thermocouple, pressure measurement and output into one

Business Terms:

Port: Beijing

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 1000 Set/Sets per Year

Payment Terms: L/C,T/T,Western Union

Contact Supplier
Temperature Calibrator HS211
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Export Package

Delivery Detail: Two weeks after received the payment

Specifications and Product Details:

Temperature Calibrator HS211 intergrates voltage, current, resistance, frequency, thermocouple, RTD, pressure measurement and output into one .It can connect external pressure module and calibrate hart instrument. That is high precision, easy operation with english graphic touch buttons. It is widely used in power,oil , chemical, metallurgy ,metering etc.

Function

Measuring and output voltage (DC V, DC mV) current (DC mA) , resistance, frequency, thermocouple, RTD.

Measuring and output S,R,B,K,N,E,J,T thermocouple .

Measuring and output Pt10,Pt100,Cu50,Cu100 RTD.

Calibrating Hart instrument directly.

Connecting digital pressure module.

Features

Intelligent operation form and instructed by graphic buttons.

It supports pressure, temperature unit switch and the mutual calculation for voltage , current and temperature values and conform to ITS90 International temperature standard .

30V voltage can be beared with any two jacks by mistaking operation.

The current terminal can be beared 1A current by mistaking operation for long time .Once the mistake operation cancelled, the items will return to the normal state.

With automatic cold junction compensation, built-in cold-junction heating module, it quickly tracks the temperature changes, and carry out cold junction sensor calibration.

Application function for simulating pressure transmitter, calibration pressure transmitter and temperature switch .

Technology specification

Using environment temperature, (-5 – 45)°C

Relative humidity, 20% - 90% No condensation

Store environment temperature, (-20 – 55)°C

Barometric pressure, (86~106)kPa

Power supply, rechargeable battery or AC 220V + 10% / 50Hz

FunctionRangeResolutionAccuracy

Measure

mV

(-10.000~80.000)mV0.1μV±(0.01%R.D+0.005%F.S)

V

(0-30.000)V0.1mV±(0.01%R.D+0.005%F.S)

mA

(0~30.000)mA0.1μA±(0.01%R.D+0.005%F.S)

Hz

(0~50.000K)Hz1Hz±(0.02%R.D+0.005%F.S)

R

R-2W(0~400.000)OHM1mOHM±(0.02%R.D+0.005%F.S)
R-3W(0~400.000)OHM1mOHM±(0.02%R.D+0.005%F.S)
R-4W(0~400.000)OHM1mOHM±(0.01%R.D+0.005%F.S)

mV

(-10.000~80.000)mV1μV±(0.02%R.D+0.005%F.S)

Output

V

(-10.000~10.000)V1mV±(0.02%R.D+0.005%F.S)

Hz

(0~50.000)KHz Amplitude 2~11V,bias±11V1Hz±(0.01%R.D+0.005%F.S)

mA

(4~20.000)mA1μA±(0.02%R.D+0.005%F.S)

OHM

(1.000~400.000)OHM10mOHM±(0.02%R.D+0.005%F.S)

DC24

---------------------±0.25V
Temperature Calibrator HS211 suppliers    Temperature Calibrator HS211 manufacturers