The TH-2001 chemiluminescence method NOx Analyzer uses modern micro single chip technique to detect, control and process signals
Port: Wuhan
Minimum Order Quantity: 1 Set/Sets
Payment Terms: ex-factory price
Contact SupplierPackaging Detail: cartons then to wooden cases
Delivery Detail: deliverying within three months after payment
Applications
Based on the gas phase chemiluminescence principle, the TH-2001 chemiluminescence method
NOx Analyzer uses modern micro single chip techniqueto detect, control and process signals. It is one of the hightech serial automatic ambient air monitoring products combined the photology,
machinery,electronics, photochemistry and modern sensor techniques and can continuously monitor
trace NO,NO2 and NOx concentrations in ambient air.
Features
Specifications
NO. | Item Name | Specifications | ||
1 | Measuring range | (0~0.5)ppm or(0~1.0)ppm(can be extended) | ||
2 | Resolution | 0.1ppb | ||
3 | Noise | Zero point noise | 1ppb | |
Span noise | 5ppb | |||
4 | Minimum detectable value | 2ppb | ||
5 | Stability | Zero drift | ±5ppb/24hr | |
Span drift | 20% Span drift | ±3ppb/24hr | ||
80% Span drift | ±5ppb/24hr | |||
6 | Precision | 20% measuring span | ±3ppb | |
80% measuring span | ±5ppb | |||
7 | Linearity | Related coefficient(r) | more than 0.999 | |
Slope rate(a) | 0.99-1.01 | |||
intercept(b) | less than ±1%F.S | |||
8 | Voltage stability | ±1%F.S | ||
9 | Response time | Rising response time(T90) | 5min(Only the analyzer itself, the gas distributing system response time not included) | |
Falling response time(T10) | 5min(see the above description) |