Home  >  Analyzers  >  Other Analysis Instruments

Semi-auto Pensky-Martens Closed Cup Flash Point Tester - LH-907

Pensky-Martens Closed Cup Flash Point Tester, -Digital display, - It has forcible wind cooling function - more info check here: http://www.buymeasuringtools.com/Semi-auto-Pensky-Martens-Closed-Cup-Flash-Point-Tester-10195885/

Quick Brief

Place of Origin: Shanghai China (Mainland) Brand Name: LH Model Number: LH-907 device: test instrument: asphalt apparatus: petroleum

Product Features:

Pensky-Martens Closed Cup Flash Point Tester

-Digital display

- It has forcible wind cooling function

Business Terms:

Port: Shanghai China

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 1000 Set/Sets per Month

Payment Terms: L/C,T/T

Contact Supplier
Semi-auto Pensky-Martens Closed Cup Flash Point Tester
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Carton

Delivery Detail: 15 days after received TT

Specifications and Product Details:

Semi-auto Pensky-Martens Closed Cup Flash Point Tester(Digital Display)

Model: LH-907

Description
LH-907 Semi-auto Pensky-Martens Closed Cup Flash Point Tester(Digital Display) is designed and made as per the National Standard GB/T261 “Test Methods for Flash Point of Petroleum Products (Closed Cup Methods)”.And refer to ASTM D93. It is used to determine the lowest temperature of a petroleum product sample when the mixture of sample vapor and air around the sample contacts and flashes when heated in a closed cup under stipulated conditions. The lowest temperature is the flash point of the sample.
Adopt microprocessor techniques and equipped with advanced software, it can control temperature, stir sample, ignite, detect, and print out test data automatically. It has a six digits display.
Feature
1. It adopts colored LCD, man-machine conversation interface, and blank buttons. It has prompts menu and guide to choose method A or B, preset temperature, sample serial number, barometric pressure, test data and other parameters.
2. It can simulate and show temperature-time curve. It has prompts for wrong operation and prompts to modify test data, test time, and other parameters.
3. It can save test data of 100 groups. It is equipped with RS-323, and 485 computer ports, so it can save more test data when it is connected to a computer. Then you can also modify some parameters.
4. It can calibrate the effect of barometric pressure to the test and calculate the modified value.
5. Differential detection. It can modify system deviation automatically.
6. It can open lid, apply flame, make detection, and print test data automatically. The test arm can be raised up and lowered down automatically.
7. It uses an electron igniter for flame ignition. It has forcible wind cooling function.
Main technical parameters
1. Operation condition
- Ambient temperature: +10°C~+30°C;
- Relative humidity: < 80%;
- Power supply: AC 220V±10%, 50±1Hz;
2. Temperature measurement
- Measurement range: ambient temperature~250°C;
- Accuracy: 0.2%;
- Resolution: 0.1 °C;
- Repeatability:
Flash Point is lower than 104 °C: error is 2 °C;
Flash Point is higher than 104 °C: error is 4 °C;
3. Basic parameters
- Interval for flame application
Flash point is lower than 104 °C: 1 °C;
Flash point is higher than 104 °C: 2 °C;
- Temperature rising rate: as per GB/T261-83;
- Stirring speed: 105 RPM
- Ignition mode: by gas flame (4 mm);
- Total power consumption: not more than 300 W;
- Dimension:
Main case: 360×110×300 (mm);
Measurement case: 320×260×200 (mm);
4. Net weight: 10 kg.

Automatic Pensky-Martens Closed Cup Flash Point Tester/Oil Tester

, PMCC Closed Cup Flash Point Tester, Closed Cup Flash Point apparatus, Flash point testing machine

Semi-auto Pensky-Martens Closed Cup Flash Point Tester suppliers    Semi-auto Pensky-Martens Closed Cup Flash Point Tester manufacturers