Salt Spray Corrosion testing machine
1.Durable quality for salt spray test;
2.Anti-corrosion, easy clean.
Port: Shenzhen
Minimum Order Quantity: 1 Set/Sets Haida
Supply Ability: 60 Set/Sets per Month Haida
Payment Terms: L/C,T/T,Western Union
Contact SupplierPackaging Detail: Standard export wooden case
Delivery Detail: 15 days after order confirmed
Salt Spray Corrosion testing machine
Salt spray corrosion testing machine adapted to the surface treatment of various materials, including paint, electroplating, organic and inorganic coating, anode processing, antirust oil and so on, after these anti-corrosive treatment to test the corrosion resistance of their products.
Specifications:
Model | HD-90 salt spray corrosion testing machine |
Inner Box Size (L×W×H) | 90×60×50 cm |
Outside Box Size(L×W×H) | 141×88×128 cm |
Equipment Material | Both inside & outside shell adopt imported P.V.C rigid plastic board, and tank cover uses imported transparent P.V.C. rigid plastic board |
Temperature Range | 35°C~55°C |
Temperature Fluctuations | ≤±0.5°C |
Temperature Uniformity | ≤±2°C |
Temperature Precision | ±1°C |
Test Chamber Temperature | Salt Spray Method (NSS ACSS)35±1°C Corrosion-resistant Testing Method(CASS)50±1°C |
Saturated air Barrel Temperature | Salt Spray Method (NSS ACSS)47±1°C Corrosion-resistant Testing Method(CASS)63±1°C |
Brine Temperature | 35°C±1°C |
Spray Quantity | 1.0~2.0 ml / 80cm2 / hr |
PH | Salt Spray Method (NSS ACSS6.5~7.2) Corrosion-resistant Testing Method(CASS)3.0~3.2 |
Lab Volume | 270L |
Brine Tank Capacity | 25L |
Power | AC220V,15A |
Multiple Safety Protection Devices | Current discharge protection, over pressure protection,over temperature protection, over load fuse protection |
Accessories | Testing Salt/ Measuring tank/ Equipment each 1 set |
Standard | CNS: 3627, 3385, 4159, 7669, 8886; JIS: D0201, H8502, H8610, K5400, Z2371; ISO: 3768, 3769, 3770; ASTM: 8117, B268 |
Salt Spray Corrosion testing machine Maintenance Procedure
1,Change the heat sink water if the test period over one month.
2,Don’t use the salt solution if unused for a week, avoid affecting the test quality.
3,Clean the inner of the chamber and release the water in heat sink if a longer time interval from the next test.
release the water from heat sink---------------open the drain value
Release the water from isolation tank--------------- pulled up the middle silicone plug.
Release the water from preheat tank--------------- Pull out the internal silicone plug.
4,ensure the test quality, please replace the glass nozzle periodically within 4500 hours.