PH, ORP, Cond, TDS, Sal, Res, DO, Temp, and so on
Minimum Order Quantity: 1 Unit/Units
Supply Ability: 100 Unit/Units per Week
Payment Terms: L/C,T/T
Contact SupplierPackaging Detail: rugged and durable cartons/wooden box
Delivery Detail: 2-3 weeks after payment receipt
I. General Introduction
v Perfect combination of advanced electron technology, sensor technology and software design.
v Can test pH, ORP, conductivity, TDS, salinity, resistivity, DO and temperature in water solution, with high accuracy and best cost performance
v Used in industrial& mine enterprises, generating station, water-treatment project, environmental protection, etc., especially used in the field
II. Characters
v 1. Multi-parameter meter, with 4 electrodes, that is, pH electrode, ORP electrode, conductivity electrode and DO electrode. After the electrode is inserted, it will be recognized automatically.
v 2. built-in microprocessor, with multiply function as automatic calibration, automatic temperature compensation, data back-up, function setup, information self-diagnosis, auto-shutoff, low-voltage display, etc.
v 3. Apply digital filtering and slip technology, improving its response speed and test accuracy intellectually. When the test data is stable, the signal is displayed.
v 4. auto-identify 13 kinds of pH standard buffer solution
v 5. pH testing models for pure water or ammoniated pure water are optional. To these two kinds of special pH testing models, both normal slope compensation and non-linearity temperature compensation to solution pH value are available, increasing its accuracy. It is suitable to be used in generating factory, petroleum chemistry, etc.
v 6. single-point calibration: adopting advanced Automatic Frequency Conversion and Automatic Voltage-Adjusting technology, the measuring range of electrode (content K=1) extends to 10 times. Only a single-point calibration can satisfies the measurement request for 0-100 mS/cm, which is the special single-point calibration function of our product.
v 7. Identify 8 kinds of conductivity standard solution automatically. Two serials can be available: the serial used by Euro and USA and the serial used by China.
v 8. Choose measuring range of conductivity automatically. To the pure water and super pure water, whose conductivity ratio is less than 10μS/cm, it can make non-linearity temperature compensation automatically, increasing its accuracy. It is suitable to be used in generating factory, micro-electronics industry, medical industry, etc.
v 9. DO electrode with temperature and salinity sensor: automatic temperature compensation, automatic salinity compensation and manual pressure compensation function.
v 10. Polar graphic DO electrode, with exclusive DO electrode calibration cap, only 3-5min to get polarized. Combined diaphragm cap for easy use. Each DO electrode is equipped of 3 spare diaphragm caps.
v 11. Use Surfaced Mounting Technology on circuit board, improving its reliability.
v 12. LCD display, with blue backlight
v 13. IP57 dust-proof and waterproof
PH | |
Range | (-2.00-19.99)pH |
Accuracy | ±0.01pH |
Automatic Calibration | (1~3) point |
ATC | (0~100) |
mV(ORP) | |
Range | ±1999mV |
Accuracy | ±0.1%FS |
Conductivity: Inside setup with function of high purified water non-linearity automatic temperature compensation | |
Range | Conductivity: (0.00~19.99)μS/cm (20.0~199.99) μS/cm (200~1999) μS/cm (2.00~19.99) μS/cm (20.0~199.9) μS/cm |
TDS:(0~100)g/L | |
Sal: (0~100)ppt | |
Resistivity: (0~100)MΩ·cm | |
Accuracy | ±1.0%FS |
Automatic Calibration | 1 point |
ATC | (0~50) |
Dissolved Oxygen | |
Range | (0~40.00)mg/L(ppm) (0~199.9)% |
Accuracy | ±0.10mg/L |
Temperature Compensation | Auto,(0~45) |
Salinity Compensation | Auto, (0~45)ppt |
Air Pressure Compensation | Manual, (66~200)KPa |
Included Electrode and Calibration Solution | 201T-S plastic pH/ATC 3-in-1 combination electrode |
301Pt-S ORP combination electrode | |
2301-S plastic conductivity electrode (ATC, K=1) | |
DO500 DO electrode 3pcs of membrane caps Inner solution for DO probe (30mL) | |
pH calibration solution(4.00, 7.00, 10.01pH/50mL) | |
Conductivity calibration solution (1413μS/cm/50mL) | |
Features of Meter | |
Data Storage | 400groups |
Automatically shutting down | 0 to 20 minutes, can be set up by user |
Display | (42 x 33)mm LCD with blue light background |
Power | 2pcs of AA batteries (1.5V x 2), life of battery >100hours |
IP rating | IP 57 dustproof and waterproof |
Size/weight | |
Meter | (65 x 120 x 31)mm/180g |
Carrying Case | (360 x 270 x 76)mm/1.7kg |