Home  >  Instrument Parts & Accessories

Platinum RTD/PT100 Elements 2.2 x 4.1mm - 2.2 x 4.1mm

PT100 Element, 1.Characterized by long-term stability, 2. precision over a broad temperature range and compatibility - more info check here: http://www.buymeasuringtools.com/Platinum-RTD-PT100-Elements-2-2-x-4-1mm-10132236/

Quick Brief

Place of Origin: Jiangsu China (Mainland) Brand Name: MICC Model Number: 2.2 x 4.1mm Usage: Industrial Theory: Temperature Sensor

Product Features:

PT100 Element

1.Characterized by long-term stability

2. precision over a broad temperature range and compatibility

Business Terms:

Port: Shanghai or Ningbo

Minimum Order Quantity: 50 Piece/Pieces

Supply Ability: 500000 Piece/Pieces per Month

Payment Terms: L/C,T/T,Western Union

Contact Supplier
Platinum RTD/PT100 Elements 2.2 x 4.1mm
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Carton

Delivery Detail: 15-60days

Specifications and Product Details:

Platinum Resistance Temperature Detector(PT100 Element):

HD 421 Pt 100-type platinum temperature sensors are characterized by long-term stability, precision over a broad temperature range and compatibility.

Main application area is the process technology.

Nominal Resistance R0Tolerance
DIN EN 60751
1996-07
Tolerance
DIN EN 60751

2009-05

Order Number

Plastic Box

100 Ohm at 0°c

Class B ( to 650°c)F 0.3 (to 650°c)

32 208 228

Class 2B (to 850°c)F 0.6 (to 850°c)

The measuring point for the nominal resistance is defined at 4mm from the end of the sensor body.

SpecificationDIN EN 60751
Temperature Range-70°c up to +850°c

Tolerance Class B: -70°c up to +650°c

Tolerance Class 2B: -70°c up to +850°c

Temperature CoefficientTC = 3850 ppm/K
LeadsPt- wire
Lead Length (L)6mm ±1mm
Long-term Tests1000h at 850°c (energized, open)

smaller then the allowed deviation according to DIN B. 1000 h

at 650°c (under current as clean MI-type) smaller then the

allowed deviation according to DIN B

Vibration Resistanceat least 40g acceleration with 8ms half sine

wave signal, depends on the installation

Shock Resistanceat least 100g acceleration at frequencies

from 10Hz up to 2000Hz, depends on the

installation

Environmental ConditionsUnhoused for dry environment only, Up to 650 in housings also

as clean MI-type possible, above 650 no reducing atmosphere,

free air admission necessary

Insulation Resistance>100 MΩ at 20°c; >2 MΩ at 650°c
Self Heating0.2 K/mW
Response TimeWater current (v= 0.4m/s): t0.5 = 0.05s

t0.9 = 0.17s

Air stream (v= 2m/s): t0.5 = 3.3s

t0.9 = 13.0s

Measuring Current20°c max. 5mA; 850°c max. 2.8mA

(self heating has to be considered)

NoteOther tolerances,values of resistance and wire lengths are

available on request.

Platinum RTD/PT100 Elements 2.2 x 4.1mm suppliers    Platinum RTD/PT100 Elements 2.2 x 4.1mm manufacturers