PH METER aquatic products industry
1.Attractive appearance,
2.High precision PH meter
3.portable PH meter
4.easy to carry
Port: guangzhou
Minimum Order Quantity: 1 Piece/Pieces
Supply Ability: 10000 Piece/Pieces per Month
Payment Terms: T/T,Western Union
Contact Supplier
PH METER aquatic products industry
PH METER Specification
*measure range: 0.00—14.00ph
*resolution:0.01ph
*accuracy:+/_0.01ph
*Automatic temperature compensation: 0°C--60°C
*Power supply: 2*1.5V(AG13 battery)
*operating temperature: 0°C--60°C
*calibration: manual at one point through trimmers
*dimension: 155mm*29mm*16mm
*weight: 48g
PH METER Operation
1. remove tne protective cap.
2. first rinse the electrode with distilled water, and suck it with filter paper.
3. turn the meter on by pressing the on key.
4. immerse the ph meter electrode in the solution to be tested.
5. stir gently and wait for the reading to stabilize.
6. after use. rise the electrode with water to minimize contamination. turn the meter on by pressing the off key.
7. always replace the protective cap after use.
PH METER Calibration
1. immerse the electrode in PH6.86(under the temperature of 25°C ) standard buffer solution of phosphate, and gently shake the electrode.
2. regulate the trimmer with a screwdriver until the buffer solution value corresponding to the measurement temperature is obtained.
3. immerse the electrode in ph 4.01 or ph 9.18 the standard buffer solution of borax.
4. after about one minute, until the buffer solution value corresponding to the measurement temperature is obtained.
Important : the instrument’s ph range must be re-calibrated whenever.
1. approximately one month has passed from the last calibration.
2. the electrode has been used in particularly taxing conditions.
3. the utmost accuracy is required.
Battery install
If the meter can not be switched on or the display fades, to change batteries, pull out the battery case and replace batteries. Pay attention to the polarity of battery.
PH METER Warranty
These instruments are warranted from all defects in material and manufacturing for a period of one year from the date of purchase.
If during this period, the repair or the replacement of parts is required where the damage is not due to negligence or erroneous operation by user, please return the parts to either repair will be effected free of charge.