1.Replacement module;
2.Memory function in case of power-down;
3.High accuracy;
4.Heating/cooling rate is 1~3 deg c each s
Minimum Order Quantity: 1 Set/Sets
Supply Ability: 500 Set/Sets per Month
Payment Terms: L/C,T/T,Western Union,MoneyGram
Contact SupplierPackaging Detail: 1.Standard exporting package;2.Customized package is also available;
Delivery Detail: 10-15days from the second working day of the deposit
Gradient PCR K960
Product combines most advanced of today’s semiconductor technology (PeltierBased), with imports of core parts, ensure that the products of amplification rate, analysis results and system reliability. Reliable guarantee for the accuracy of the Temperature extended control mode which is closer to required experiment temperature control and is able to effectively avoid the system error caused by the disaccord of the temperature points among the instrument’s display temperature, actual block temperature and the temperature required for reagents. So as to improve the accuracy of the experiment and ensure the high efficiency. Strict temperature control debugging program makes sure that each instrument can meet the needs of different experiment.
12 channel temperature probes detect simultaneously, which ensure the homogeneity of sample temperature. The hermetic-space technique can efficiently eliminate PCR margin reaction. The technique of outside temperature probe tracing the inside curve testing can effectively ensure the accuracy of sample temperature.
Block types
The sample of replacement
Features
replacement mode.
Large size super-high-definition LCD screen.
Intuitive, friendly user interface makes program easier and faster.
Memory function in case of power-down.
Solemn, elegant appearance, innovative model.
Unique rotating stall heat-regulating function.
Optimal panel design for human ,more convenient operation.
Heat lid could be stopped at any angle.
Handlemodule, more secure and convenient module replacement, improving
using efficiency and expanding using years.
The sample temperature control
Instrument working condition
Ambient temperature:5°C-30°C
Relative humidity:90% less
Power supply:AC110V±22V,220V±22V,250VA,50Hz±1Hz
Instrument storage condition
Ambient temperature: -20°C-55°C
Relative humidity: 90% less
Interface of program displays
A 0.2ml×96
B 0.5ml×54
C 0.2ml×96+0.5ml×77
D 384 WELL
Performance index
Model | K960-A,K960-B,K960-C,K960-D |
Capacity | 96x0.2mL(A),54x0.5mL(B),96x0.2mL+77×0.5ml(C),384well(D) |
Temp range | 0°C—99.9°C(Rt≤30°C) |
Max Heating rate | ≥4.0°C/s |
Max cooling rate | ≥4.0°C/s |
Heating/cooling rate | 1.0°C/s—3.0°C/s( adjustable) |
Uniformity | ≤±0.2°C |
Accuracy | ≤±0.1°C |
Gradient temp range | 30°C—99°C |
Gradient spread | 1°C—30°C |
Gradient Uniformity | ≤0.2°C |
Heat lid temp | 30°C—110°C |
Environment temperature identification | manual identification |
Temp control | block,tube,calculated |
Stored program no | 200 |
Max.No .of Cylce | 99 |
Display | 5.7inch,320*240pels LCM |
Communication | USB2.0 Rs232 |
Size | 380mm(L)*270mm(W)*250mm(H) |
Weight | 7.2kg |
Universal voltage International. 110V-220V |