Model GC112A Gas Chromatograph. - Model GC112A Gas Chromatograph

Model GC112A Gas Chromatograph:Self-diagnosis system, power-fail protection, file storage and call, limited temperature setting. - more info check here: http://www.buymeasuringtools.com/Model-GC112A-Gas-Chromatograph-10179330/

Quick Brief

Place of Origin: China (Mainland) Brand Name: SPSIC. Model Number: Model GC112A Gas Chromatograph

Product Features:

Model GC112A Gas Chromatograph:Self-diagnosis system, power-fail protection, file storage and call, limited temperature setting.

Business Terms:

Payment Terms: T/T

Contact Supplier
Model GC112A Gas Chromatograph.
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Specifications and Product Details:

Model GC112A Gas Chromatograph

Characteristics

1. Brand-new designed microcomputer temperature control system of world advanced control technology. High accuracy of temperature control (optimum: ±0.05°C), high reliability and anti-interference.

2. Control and operating parameters (including for detectors) set up by the keyboard. Self-diagnosis system, power-fail protection, file storage and call, limited temperature setting. Accurate display of setting and actual value of temperature control of each path, bridge current of TCD, sensitivity of FID, retention and analysis time.

3. Four paths of independent control sections: column oven, sample injector, detector and thermal conductive pond.

4. 5-ramp programmed temperature rising control. Auto-drop of temperature of column oven (auto-opening the door behind the oven) to a fast cooling.

5. Dual FID provided with options of GC112A-TCD, GC112A-ECD, GC112A-FPD and GC112A-NPD, gas sampling valve, transformation oven, absorber and pyrolysis absorber available for selection. Two detectors can be installed at the same time with high sensitivity. Short time to be stabilized and easy to clean and install nozzle.

6. Packed column and capillary column (0.53 large diameter or small diameter) can be installed easily including a complete set of split sampling connector and tailing connector to be installed free of charge. Optional 0.53mm large diameter capillary direct sampler is available for choice.

7. Analysis can be conducted in packed or capillary column with multiple sampling methods.

8. Packed column: on-column sampling, instantaneous vaporized sampling and gas sampling.

9. Capillary column: split sampling and 0.53mm large diameter capillary direct sampling.

10. Construction of dual gas path, dual sample injector and dual packed column system can effectively avoid baseline undulation and drift in real time and facilitate programmed temperature rising analysis. Accurate and scaled type gas path control valves will provide a high repeatability and stability.

11. Advanced functions: auto ON/OFF temperature control system, permanent data storage, file storage and call and over-temperature protection.

12. Large volume column (300×280×270) can hold capillary column and dual packed column. Built-in heating filament structure is in favor of quartz capillary analysis.

13. A complete set of accessories including purifier, pressure-reducing valve, gas path tube, wrench and tools, injection needle and connectors of all kinds are provided free of charge and are ready to connect with supply-air system prepared by user.

Specifications

Control Range7°C~400°C above room temperature. (increment of 1°C)
Control ObjectsColumn oven, detector, thermal conductive pond and sample injector
Programmed Temperature Rising5 ramps
Rate of Programmed Temperature Rising0.1°C~40°C (increment of 0.1°C)
Thermostatic Time0~655min (increment of 1min)
Sensitivity (minimum testability)M≤5×10-11g/s (sample: C16)
Optimum Testing ResultM≤1×10-11g/s
Noise≤5×10-14A
Drift≤6×10-13A/h
Linear Range≥106
Sensitivity≥2000mV·ml/mg
Noise≤20μV
Drift≤30μV
Linear Range≥104
Sensitivity (maximum testability)≤2×10-11g(P)/s (sample P in the parathion methyl)

Optimum testing result ≤3.8×10-12g(P)/s.

Sensitivity (maximum testability)≤1×10-10g(S)/s (Sample S in the parathion methyl)

Optimum testing result ≤2.7×10-11g(S)/s.

Linear Range≥103 (sample P parathion)

≥102 (sample S parathion)

Stability≤30μV/h
Noise≤10μV
Starting TimeNot greater than 2 hours
Sensitivity (maximum testability)Dt(N) ≤5pg(N)/s

Optimum testing result 0.5ppm

Sensitivity (maximum testability)Dt(P) ≤0.5pg(P)/s

Optimum testing result 0.2ppm

Drift≤8×10-13A/h
Noise≤5×10-13A/h
Linear Range≥103
Max Application Temperature360°C

Optional Accessories

1. TCD

2. FPD

3. NPD

4. CDP-4A data processor

5. Software with I/O port

Overall Dimension & Weight

Shipping Package Dimension(990×630×690)mm
Net Weight50kg
Gross Weight75kg