Home  >  Pressure Measuring Instruments  >  Pressure Transmitters

Microcyber Liquid Level Pressure Transmitter - NCS-PT105II

1.support FF,PA,Hart protocal, 2.0.075%high accuracy, 3.4-20mA,digital signal multi output, 4.measure gas,liquid and steam,etc - more info check here: http://www.buymeasuringtools.com/Microcyber-Liquid-Level-Pressure-Transmitter-10330072/

Quick Brief

Place of Origin: Liaoning China (Mainland) Brand Name: Microcyber Model Number: NCS-PT105II protocal: FF, Hart, PA accuracy: ±0.075% multi output: 4-20mA, digital signal

Product Features:

1.support FF,PA,Hart protocal

2.0.075%high accuracy

3.4-20mA,digital signal multi output

4.measure gas,liquid and steam,etc

Business Terms:

Port: Shenyang/China

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 10000 Set/Sets per Year

Payment Terms: T/T

Contact Supplier
Microcyber Liquid Level Pressure Transmitter
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: normal cartons

Delivery Detail: within only 1 day to 2 months according to the order amount

Specifications and Product Details:

1. Profile

NCS-PT105Ⅱ series Smart Pressure Transmitter with advanced, mature, reliable 3151
capacitance sensors has been designed meticulously by combining advanced
microprocessor technology and digital capacitance measurement technology.
The powerful functions and high-speed computing capability of the
microprocessor make it have excellent qualifications such as smart, high
precision, high reliability, stable zero and so on. Its LCD can display many physical
parameters (e.g. pressure, temperature, current and so on). It can realize the
functions such as zero adjustment, range settings by key-press operation, and
it is easy for field testing.


NCS-PT105Ⅱ series Smart Pressure Transmitter supports HART, FF, and PA protocol and
can measure pressure, differential pressure, liquid level, flow, and other
industrial parameters. It can be widely used in the petroleum, chemicals,
electricity, and metallurgical industries, etc.

2. Pictures

3. Parameters

Number


Display


Reason


Methods


1


Output
current value is 0


a. power
failure


b. wire
cutting


a. modify
power


b. check
wire


2


HART
smart transmitter output current value exceeds limit


Failure
connection between the


sensor
and the circuit board.


Check the
connection.


3


HART
smart transmitter current fixed on


4mA,add pressure the output,no change.


a.
instrument in multi-point mode


b.
leakage


a. modify
slaves address in single mode


b. check
the connection of gas pipe


4


Smart
transmitter communication failure


a.
connection failure


b.
multi-mode


a. check
the wire loop


b. check
net

Microcyber Liquid Level Pressure Transmitter suppliers    Microcyber Liquid Level Pressure Transmitter manufacturers