IEC156 Insulating Oil Detector,BDV Breakdown Voltage Testing:
1.Voltage test precision 15.
2.Breakdown sensitivity 2KV.
Port: Chongqing
Minimum Order Quantity: 1 Set/Sets no
Supply Ability: 100 Set/Sets per Year no
Payment Terms: L/C,T/T
Contact SupplierPackaging Detail: wooden case
Delivery Detail: 7days
IEC156 Insulating Oil Detector:
IIJ-II automatic oil tester is special used in the electric power industry for measurement for breakdown voltage of insulation oil. The dielectric strength is a comprehensive index for transformer oil,IIJ-II oil tester is designed according to international standard IEC156 dielectric strength. It adopts built-in printer, LCD displayer, industrial single chip as controller, new type I/O interface, apply large integrate circuit etc. This oil tester has high capability of anti-jamming and can run in open air. The system clock can be run when be power off
Features:
1. Small size, light weight, easy operations
2.Adopt menu management, impute parameter is ok
3.Can be stored ninety-nine kinds of oil test data which is convenient to print and adjust the data.The test data can be stored 10 years when the electrical is power off
4.The system clock can be run when be power off
5.High capability of anti-jamming, high precision and can run in open air.
Technical Parameters:
Power supply | AC220V±10%,50Hz |
Output voltage | 0~100KV(60KV,80KV,100KV) |
Capacity | 1.2KVA(1.6KVA, 2.0KVA) |
Speed of pressure rise | about 2KV/S |
Voltage test precision | ±1% |
Breakdown sensitivity | <2KV |
Wave difference | ≤3% |
Time of breakdown | ≤10ms |
Working condition- temperature | 0°C~40°C |
- humidity | The biggest comparatively humidity85% |
Reserve circumstance- temperature | -20°C~60°C |
- humidity | The biggest comparatively humidity75% |
- Working altitude | <1500m (if over 1500m can be designed specially) |
Dimension | 480mm*460mm*460mm |