High sensitivity green PH Meter - ph-c

High sensitivity green PH Meter, 1.Large-screen LCD display ,easy viewing durable, 2.high precision, 3.Analysis machine - more info check here: http://www.buymeasuringtools.com/High-sensitivity-green-PH-Meter-10122624/

Quick Brief

Place of Origin: Guangdong China (Mainland) Brand Name: 3win Model Number: ph-c color: green

Product Features:

High sensitivity green PH Meter

1.Large-screen LCD display ,easy viewing durable

2.high precision

3.Analysis machine

Business Terms:

Port: guangzhou

Minimum Order Quantity: 100 Piece/Pieces

Supply Ability: 10000 Piece/Pieces per Month

Payment Terms: T/T,Western Union

Contact Supplier
High sensitivity green PH Meter
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Specifications and Product Details:

High sensitivity green PH Meter

1.Large-screen LCD display( for easy viewing durable)
2.high precision
3.pen shape Measurement & Analysis Instruments used for PH

SPECIFICATIONS

Range : 0.00to 14.00 pH

Resolution : 0.01 pH

Accuracy : ±0.05 pH

Batteries : 4 x 1.5V alkaline

ATC: 0°C to 60°C

Environment : RH 95% max; 0 to 60°C

Dimensions : 172x26x26mm

Weight : 73 g

OPERATION

·Remove the protective cap. Do not be alarmed if white crystals

appear around the cap. This is normal with pH electrodes and

they dissolve when rinsed with water.

·Turn the tester ON by pressing the 'ON/OFF' button.

·Immerse into solution up to the max immersion level.

·Stir gently and wait until the display stabilizes.

·After use, Turn the tester off by pressing the ON/OFF button,

use the water clean the electrode and replace the protective cap.

Do not use distilled or deionized water for storage purposes.

·Large fluctuations in readings could be due to lack of calibration,

dry electrode or rundown batteries.

CALIBRATION

·Immerse the tester up to the maximum immersion level in pH 6.86 buffer solution.

·Allow the reading to stabilize and using a small screwdriver adjust the calibration trimmer to read 6.86

·Immerse the electrode in PH4.01 or PH9.18 the standard buffer solution of borax. Until the display

shows 3.96-4.06PH or 9.13-9.23PH.

Always use fresh buffers for calibration.

Important The instrument must be re-calibrated whenever.

·The electrode or battery has been replaced.

·Approximately one mouth has passed from the last calibration.

·The electrode has been used in particularly taxing conditions.

·The utmost accuracy is required.

BATTERY REPLACEMENT

When the battery symbol -----appears on the display, this indicates a low battery and only 1 hour

of continuous use remain. though the unit may continue to function ,the accuracy of the unit will be

affected beyond the 1 hour.twist open the battery compartment lid. remove old batteries and replace

with fresh ones noting polarity as show in battery compartment.

ELECTRODE REPLACEMENT

1.Remove the protective cap.

2.Unscrew the plastic ring on the top of the electrode .

3.Pull out the electrode and replace it with a new one. make sure the gaskets are in place before screwing

back the ring.

High sensitivity green PH Meter