1.Certificate: ISO 9001;
2.Accuracy class: 0.02
3. Full english interface
Port: As per your request
Minimum Order Quantity: 1 Set/Sets
Supply Ability: 3 Set/Sets per Week
Payment Terms: T/T
Contact SupplierPackaging Detail: Plywood case
Delivery Detail: 30 days
High Precision Meter reference device
1. High precision : Class 0.02 for measurement of U,I,P,E
2. Multi-function: measure voltage,current, phasor, frequency,Power, atctive/reactive energy, errors at the same time.
3. Wide measuring range: 30V~400V for Voltage,.1A~100A for current
4. convenient operation: equipped with ARM series high performance CPU as the main controlling unit, operating by Windows CE.NET and with 640X480TFT true color, high light LCD.
5. Build-in testing plan as per energy meter calibration procedure, autosaving testing data
6. Fully digital product: adopping high speed A/D +DSP technology
7. Multi additional functions, include:
a. Symmetery calculation for Voltage, Current and Phasor
b. Stability calculation for voltage, current and phasor
c. Interaction calculation for volatge, current and phasor
d: wave display, harmonic analysis, distortion calculation
e: Calibrate Voltmeter, amperemeter,phasor indicator, frequency indicator and power indicator
f: volage calculation for the same end
g: intensity mesurement for magnetic field
h. Errors, S value measurement and multi-output consistency and short term stability variation measurement.
i: exterior interface: RS232, USB convenient for data interchange, display and output.
SMS PC software
1- Design according to JJG597-2005. automatically make conclusion. Auto- Data integer
2- With 1024*768 interface, accord with basis dispaly requirement of Laptop
3- Print model and modification of printing certification can be modified in WORD format
4- Realize all testing items according to related standard
Voltage measuring range: | 3×60,3×100,3×200,3×400V auto/manual -shift | |
Current measuring range: | 0.1A, .25A,0.5A,1A,2.5A,5A,10A,20A,40A100A Auto/manual shift | |
Accuracy: | Voltage | 0.02% |
current | 0.02% | |
Basic Errors | 0.02% | |
Frequency | ±0.01Hz | |
Active power | 0.02% | |
Reactive power | 0.1% | |
Reactive energy | 0.1% | |
Phase angle | ±0.05° | |
Power supply | AC220±10% | |
Power consumption | <30VA | |
Phase angle measuring range | 0°--359.999° resolution:0.001° | |
Frequency scope | 45.0000Hz---65.0000Hz resolution:0.0001Hz | |
24Hour variation | <0.004% | |
yearly variation | <0.01% | |
Dimension | 540*486*182mm | |
Working temperature | 20°C±10°C | |
Relative humidity: | 25%--90% no condensation | |
connect mode | 4 wire active 4 wire reactive true 4 wire reactive cc 3 wire active 3 wire reactive true 3 wire 2-element reactive cc 3 wire 3-element reactive cc 3 wire man-made netual reactive 3 wire 60° reactive |