*The Multifunction Process Calibrator is a powerful, yet easy-to-use field calibrator
*Loop Power Voltage: 24V
*Accuracy: 10
Port: FOB shenzhen or guanzhou
Minimum Order Quantity: 2 Piece/Pieces
Supply Ability: 1000 Piece/Pieces per Day
Payment Terms: T/T,Western Union,paypal
Contact SupplierPackaging Detail: 1pcs/color box/ 20pcx/CTN.
Delivery Detail: 5-7day
FEATURES
In addition to the functions listed in Table 1,the calibrator also has the following characteristics:
1.A split-screen display. The upper display allows users to measure volts, current, and pressure only.The lower display allows the user to measure and source volts, current, pressure, resistance temperature detectors, thermocouples, frequency,and ohms.
2.A thermocouple (TC) input/output terminal with cold junction temperature compensation,ambient temperature can be set and calibrate within the menu function.
3.Each parameters can manual and automatic stepping and ramping output, signal output can be arbitrarily set within the 0% and 100% of the value. (In addition to the pressure output)
4.In the configuration menu can be set manually : Power off time for Instrument Backlight off time Screen Contrast Cold junction compensation ON/OFFHART resistence ON/OFF and ETC. You can also set the output amplitude and duty cycle for frequency manually in configuratiion menu.
5.Have calendar clock function,it can be calibrated in configuration menu.
6.It can store 8 sets of measurement and output data and call the showing at any time,all the settings and the functions of the last operation be auto-stored after power off.
Measure Accuracy: | |||
Function | Measure Singal | Measure Accuracy | Output |
0.000—30.000 V | 0.02%+3 counts (The upper display) | ||
0.000—24.000 V | 0.02%+ 3 counts (The lower display) | ||
0.000—90.000 mV | 0.01%+ 10 counts | ||
DC Voltage | -10.000 —80.000mV 0.01 % + 10 counts (use the thermocouple jack ) | ||
DC Current | 0.000—24.000 mA | 0.02%+ 4 counts(The upper display) | |
0.000—24.000 mA | 0.02%+ 4 counts(The lower display) | ||
Resistance | 0.00— 400.00Ω | 0.1Ω (4 wire), 0.15Ω (2 wire and 3 wire) | |
400.0—4000.0Ω | 0.5Ω (4 wire), 1Ω (2 wire and 3 wire) | ||
Frequency | 0.010—20.000 kHz | ±4Hz | |
0.001—39.999.0Hz | ±0.01Hz | ||
40.00 —99.99 Hz | ±0.05Hz | ||
100.0 —500.0 Hz | ±0.5Hz | ||
Pressure | Depends on the pressure module | Depends on the pressure module | |
Output Accuracy: | |||
Output | Range | Accuracy | |
DC Voltage | 0.000— 20.000 V | 0.02%+3 counts | |
0.000—90.000 m V | 0.01%+10 counts | ||
-10.000 mV— 80.000 mV 0.01 % + 10 counts (use the thermocouple jack ) | |||
DC Current | 0.000—24.000 mA | 0.02%+ 4 counts | |
5.00 — 400.00 Ω | 0.15 Ω (Excitation current 0.15 to 0.5 mA), 0.1Ω (Excitation current 0.5 to 2 mA) | ||
Resistance | 400.0 —4000.0 Ω | 0.5Ω(Excitation current 0.05 to 0.8 mA) | |
Frequency | 0.040—2.000 kHz | ±2Hz | |
2.000—4.000 kHz | ±8Hz | ||
4.000—8.000 kHz | ±32Hz | ||
8.000—10.000KHz | ±50Hz | ||
Output Hz | 0.030—99.999Hz | ±0.005Hz | |
Pressure | Depends on the pressure module | Depends on the pressure module |
RTD and Thermocouple: | ||
Type | Measure Accuracy | Output Accuracy |
PT-100 (385) | 0.3°C(4 wire), 0.5 °C(2 wire and 3 wire) | 0.3 °C |
PT-100 (3926) | 0.3°C(4 wire), 0.5 °C(2 wire and 3 wire) | 0.3 °C |
PT-100 (JIS) | 0.3°C(4 wire), 0.5 °C(2 wire and 3 wire) | 0.3 °C |
PT-200 (385) | 0.8°C(4 wire), 1.6 °C(2 wire and 3 wire) | 0.8 °C |
PT-500 (385) | 0.4°C(4 wire), 0.9 °C(2 wire and 3 wire) | 0.4 °C |
PT-1000 (385) | 0.2°C(4 wire), 0.3 °C(2 wire and 3 wire) | 0.3 °C |
Cu50 | 0.4°C(4 wire), 0.6 °C(2 wire and 3 wire) | 0.4 °C |
Cu100 | 0.4°C(4 wire), 0.6 °C(2 wire and 3 wire) | 0.4 °C |
J: | 0.5 °C | 0.7 °C |
K: | 0.8 °C | 0.8 °C |
T: | 0.5 °C | 0.5 °C |
E: | 0.5 °C | 0.5 °C |
R: | 1.5 °C | 1.0 °C |
S: | 1.5 °C | 1.5 °C |
B: | 1.8 °C | 1.0 °C |
N: | 0.8°C | 0.8 °C |
Loop Power Voltage: 24V
Accuracy: 10%
Maximum Current: 24 mA, Short-circuit protection