Place of Origin: Chongqing China (Mainland) Brand Name: gold Model Number: GD-264A Acid Number Tester Power: Electronic Usage: Acid Number Tester usge: acid number tester: standard : ASTMD664

Product Features:

The Acid Number Tester is suitable to determine acid number of transformer oil and turbine oil.

Business Terms:

Port: chongqing

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 100 Set/Sets per Week

Payment Terms: T/T

Contact Supplier
GD-264A Automatic transformer oil Acid Number Tester /ASTM D 664
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: wooden case

Delivery Detail: 7 dasy

Specifications and Product Details:

Payment & Shipping Terms

Supply Capacity

Payment term :T/T or L/CProduction capacity :500pcs/month
MOQ:1 pcPacking :Wooden case
PricenegotiableDelivery wayBy Air
Means of transport :FOB , CIF , CFRDelivery time :10 days
GD-264A Acid Number Tester of transformer oil and turbine oil
Purpose
The Acid Number Tester is suitable to determine acid number of transformer oil and turbine oil.
It is designed and made as per the National Standard GB/T264-83 “Test Methods for Acid Number of Petroleum Products”. It takes a blue LCD for display. It can make determination, print test data automatically using a microprocessor. Test procedures are automatically controlled by the microprocessor. Operation procedures are intuitive and simple. Adopt a magnetic stirrer, it is safe, reliable, delicate, and there is less noise during test procedure.
Technical parameters
1. Power supply: 220V±10%, 50Hz;
2. Power consumption: less tahn 100W;
3. Measurement range: 0.001~0.999 mg KOH/g;
4. Resolution: 0.001 mg KOH/g;
5. Test accuracy:
(1) Acid number 0.1: ±0.005 mg KOH/g;
(2) Acid number ≥0.1: 5% of displayed value;
6. Repeatability: 0.004 mg KOH/g;
7. Ambient temperature: 10°C~40°C;
8. Relative humidity: 75%;