Home  >  Analyzers  >  Other Analysis Instruments

GD-261B Semi-auto Pensky-Martens Closed Cup Flash Point Analyzer (Digital Display) - GD-261B

It is used to determine the lowest temperature of a petroleum product sample when the mixture of sample vapor and air. - more info check here: http://www.buymeasuringtools.com/GD-261B-Semi-auto-Pensky-Martens-Closed-Cup-Flash-Point-Analyzer-Digital-Display-10227495/

Quick Brief

Place of Origin: Shanghai China (Mainland) Brand Name: Gold Model Number: GD-261B Function:: Flash point analyzer Specimen:: petroleum products Method:: Pensky-Martens Closed Cup Methods Warranty:: 1 Year

Product Features:

It is used to determine the lowest temperature of a petroleum product sample when the mixture of sample vapor and air.

Business Terms:

Port: Shanghai

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 100 Set/Sets per Month

Payment Terms: T/T

Contact Supplier
GD-261B Semi-auto Pensky-Martens Closed Cup Flash Point Analyzer (Digital Display)
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Wooden Case

Delivery Detail: 10 days

Specifications and Product Details:

GD-261B Semi-auto Pensky-Martens Closed Cup Flash Point Analyzer (Digital Display)

Purpose

The instrument is designed and made as per the National Standard GB/T261 “Test Methods for Flash Point of Petroleum Products (Closed Cup Methods)”. It is used to determine the lowest temperature of a petroleum product sample when the mixture of sample vapor and air around the sample contacts and flashes when heated in a closed cup under stipulated conditions. The lowest temperature is the flash point of the sample.

Adopt microprocessor techniques and equipped with advanced software, it can control temperature, stir sample, ignite, detect, and print out test data automatically. It has a six digits display.

Main parameters
(I) Operation condition
1. Ambient temperature: +10°C~+30°C;
2. Relative humidity: < 80%;
3. Power supply: AC 220V±10%, 50±1Hz;
(II) Temperature measurement
1. Measurement range: ambient temperature~250°C;
2. Accuracy: 0.2%;
3. Resolution: 0.1 °C;
4. Repeatability:
(1) Flash Point is lower than 104 °C: error is 2 °C;
(2) Flash Point is higher than 104 °C: error is 4 °C;
(III) Basic parameters
1. Interval for flame application
(1) Flash point is lower than 104 °C: 1 °C;
(2) Flash point is higher than 104 °C: 2 °C;
2. Temperature rising rate: as per GB/T261-83;
3. Stirring speed: 105 RPM
4. Ignition mode: by gas flame (4 mm);
5. Total power consumption: not more than 300 W;
6. Dimension:
(1) Main case: 360×110×300 (mm);
(2) Measurement case: 320×260×200 (mm);
7. Net weight: 10 kg;
GD-261B Semi-auto Pensky-Martens Closed Cup Flash Point Analyzer (Digital Display) suppliers    GD-261B Semi-auto Pensky-Martens Closed Cup Flash Point Analyzer (Digital Display) manufacturers