Home  >  Analyzers  >  Other Analysis Instruments

GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93 - GD-261 PMCC Flash Point Tester

GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93 is suitable to determine closed cup flash point of petrol - more info check here: http://www.buymeasuringtools.com/GD-261-PMCC-Flash-Point-Tester-closed-cup-flash-point-tester-ASTM-D-93-10271916/

Quick Brief

Place of Origin: Chongqing China (Mainland) Brand Name: GOLD Model Number: GD-261 PMCC Flash Point Tester

Product Features:

GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93 is suitable to determine closed cup flash point of petrol

Business Terms:

Port: shanghai

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 1000 Set/Sets per Month

Payment Terms: T/T

Contact Supplier
GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: wooden case

Delivery Detail: 7 day

Specifications and Product Details:

GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93

Purpose

The closed cup flash point tester is designed and made as per the National Standard of People’s Republic of China GB/T261-2008 “Test Methods for Flash Point of Petroleum Products (Pensky-Martens Closed Cup Methods)”. It is suitable to determine closed cup flash point of petroleum products having a flash point higher than 40 °C as per GB/T261-2008.
Main technical specification and parameters
1. Power supply: AC 220V±10%, 50 Hz;
2. Heating device:
(1) The furnace is made of SiC materials; its power is 600 W;
(2) Heating power is adjustable in the range of 0~600W;
3. Sample temperature rising rate: 1~12°C per minute, manually controlled;
4. Electric stirring device:
(1) Stirring motor: 45TYZ Motor;
(2) Driving device: by a soft shaft;
(3) Stirring blade specification: 8×40 mm;
5. Stirring speed: As per the requirement of GB/T261-2008;
(1) 90~120 RPM for procedure A;
(2) 250±10 RPM for procedure B;
6. Standard oil cup: As per the requirement of GB/T261-2008;
(1) Inner diameter: 50.7~50.8 mm;
(2) Depth: 55.7~56.0 mm;
(3) Depth of Marked line on the oil cup: 33.9~34.3 mm;
(4) Oil sample cubage: about 70 ml;
7. Ignition device (Flame application device):
(1) Ignition source: Coal gas (or other civilian combustible gas);
(2) Aperture of ignition device: 0.8 mm;
8. Thermometer: Rod type Mercury Thermometer, as per GB/T261-2008;
(1) Temperature range: -5~110°C, Each scale division is 0.5 °C;
(2) Temperature range: 20~150°C, Each scale division is 1.0 °C;
(3) Temperature range: 90~370°C, Each scale division is 2 °C;
9. Ambient temperature: ≤35°C;
10. Relative humidity: ≤85%;
11. Total power consumption: not more than 650 W;
GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93 suppliers    GD-261 PMCC Flash Point Tester/closed cup flash point tester/ASTM D 93 manufacturers