Purpose
The Automatic Multifunctional Degassing Oscillation Tester is designed and made as per the National Standard GB/T17623 Test Methods for Componental Contents of Gases Dissolved in Insulating Oil by Gas Chromatography Method.GD-17623 Degassing Oscillation Tester is suitable to separate gas dissolved in the insulating oil by oscillation methods.
Technical specifications
1. Display: By a large LCD;
2. Temperature range: room temperature99.9;
3. Temperature controlling accuracy:
(1) Room temperature50: ±0.2;
(2) 5080: ±0.3;
4. Oscillation frequency: 275±3 times/min;
5. Oscillation amplitude: 35 mm;
6. Controlling time:
(1) Oscillate for 20 min, still it for 10 min and keep constant at 50 ;
(2) Oscillate for 5 min, still it and make an alarm. Keep constant at 70 ;
(3) Set oscillation time and still time as your requirement from 0 to 99 min;
(4) Set temperature from 0 to 99 ;
7. Sample number: 8 pieces of syringe of 100 ml;
8. Power: 1000 VA;