Home  >  Other Measuring & Analysing Instruments

GD-17040 X-ray Fluorescence Sulfur Oil Analyzer - GD-17040

X-ray Fluorescence Sulfur Oil Analyzer, The instrument is specially designed to determine sulfur content in the petroleums - more info check here: http://www.buymeasuringtools.com/GD-17040-X-ray-Fluorescence-Sulfur-Oil-Analyzer-10227823/

Quick Brief

Place of Origin: Shanghai China (Mainland) Brand Name: Gold Model Number: GD-17040 Standard:: ASTM D4294 Warranty:: 1 Year

Product Features:

X-ray Fluorescence Sulfur Oil Analyzer

The instrument is specially designed to determine sulfur content in the petroleums

Business Terms:

Port: Shanghai

Minimum Order Quantity: 1 Set/Sets

Supply Ability: 100 Set/Sets per Month

Payment Terms: T/T

Contact Supplier
GD-17040 X-ray Fluorescence Sulfur Oil Analyzer
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Wooden Case

Delivery Detail: 10 days after receive payment

Specifications and Product Details:

GD-17040 X-ray Fluorescence Sulfur Oil Analyzer

Purpose
Environment pollution is attracting more and more governments and public concern. The cry for protection to ecological environment and atmospheric environment is increasingly louder. The key culprit to atmospheric destruction and pollution is discharge of CO2and SO2, which are from combustion of power fuels. Lowering down the sulfur content in the gasoline, diesel oil, heavy oils used in vehicles, ships, and airplanes will highly reduce pollution to atmosphere. Europe and our government have promulgated a series of standards on sulfur content of petroleum products to restrict its production and discharge. SO2may also cause catalyst poisoning in petrochemical industry and form acid rain in the atmosphere.
The instrument is specially designed to determine sulfur content in the petroleum products. Adopts energy-dispersive principle, and with integrated design, it is rapid and accurate. It can meet the requirements of National Standard GB/T 17040 and GB/T11140 on repeatability and reproducibility, and it can also meet requirements of ASTM D4294-03 on these indexes. It provides a measure to determine sulfur content during petroleum or petrochemical production process.
Main technical parameters
1. Measurement range: 7 ppm to 5%;
2. Precision:
(1) Repeatability: <0.029 ( S+0.6);
(2) Reproducibility: <0.063 ( S+0.6);
3. Sample quantity: 2 to 3 ml (it is equal to sample depth of 3 mm to 4 mm);
4. Measurement time: 60, 120, 240, 300, and 600 s at random;
5. It can make determination automatically for single sample. Measurement times: can set 2, 3, 5, 10, and 50 times at random; it will show average value and standard deviation at end of measurement.
6. Calibration curve numbers: it can save 9 calibration curves. 5 pieces of them are linear equation in one unknown and 4 pieces of them are binomial parabola.
7. Working condition:
(1) Temperature: 5 to 40 °C;
(2) Relative humidity: ≤85% (30 °C);
(3) Power supply: AC 220V±20V, 50 Hz;
(4) Rated power: 30 W;
8. Size and weight: 468 mm×368 mm×136 mm; 13 kg;

<

GD-17040 X-ray Fluorescence Sulfur Oil Analyzer suppliers    GD-17040 X-ray Fluorescence Sulfur Oil Analyzer manufacturers