Electronic temperature measurement, electronically automative ignition, digital display, integrated open and closed cup
Port: Dalian
Minimum Order Quantity: 1 Set/Sets
Supply Ability: 10 Set/Sets per Week
Payment Terms: T/T
Contact SupplierPackaging Detail: cartons
Delivery Detail: within one week after receiving T/T in advance
Function Standard
Electronic temperature measurement, electronically automative ignition, digital display, integrated open and closed cup, safe and reliable; closed cup flash point range: 30~260oC, open cup flash point range: 102~280oC
Technical Parameters
Test flash point temperature range: closed-cup flash point: 30~260oC
open-cup flash point: 102~280oC
Recorder allowed absolute inaccuracy range:
closed-cup flash point: 30~104 (including 104oC) ±2oC
Beyond 104oC ±5oC
open-cup flash point: ±5oC
Measuring result astringency(repetitive rate):
closed-cup flash point: 30~104oC (including 104oC) 2oC
Beyond 104oC 5oC
open-cup flash point: 5oC
Measuring result repeatability:
closed-cup flash point: 30~104oC (including 104oC) 4oC
Beyond 104oC 8oC
open-cup flash point: 16oC
Model heating speed before flash point: 5~6/min
Appear electric spark exoelectric discontinuity,second:
closed-cup flash point: 30~104oC (including 104oC)≤10oC
Beyond 104oC ≤24oC
Measurement: 300×260×180mm (L×W×H)
Weight: Not more than 10kg
Power: Not more than 70VA
Voltage: 220+10%-15%
Frequency: 50±1Hz