Home  >  Analyzers  >  Other Analysis Instruments

BT-1600 Image Particle Size Analyzer - GA-1600

Sample Adding Method:Static slide, Auto cut, auto calibration, Multi-image process - more info check here: http://www.buymeasuringtools.com/BT-1600-Image-Particle-Size-Analyzer-10259463/

Quick Brief

Place of Origin: Guangdong China (Mainland) Brand Name: BEATOP Model Number: GA-1600 Measuring Range: 1-3000μm Max magnifying power: 4000X(optical + electric ... Max Resolution: 0.1μm Repeatability: <3% (CRM D50) Particle size distribution type: volume, area, ... Output Items: Results output: Original paramete...

Product Features:

Sample Adding Method:Static slide

Auto cut, auto calibration

Multi-image process

Business Terms:

Port: Zhuhai

Minimum Order Quantity: 1 Set/Sets n/a

Supply Ability: 150 Set/Sets per Month n/a

Payment Terms: L/C,T/T

Contact Supplier
BT-1600 Image Particle Size Analyzer
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Seaworthy packing

Delivery Detail: 1 month after order confirmation

Specifications and Product Details:

GA-1600 image particle analysis system is multi-functional including image distribution test and particle shape analysis. The integration of the traditional microscopy and advanced image processing technology, it consists of optical microscope, CCD camera, image acquisition system, computer, printer, etc. The basic operation is performed by firstly taking the picture of the image in microscope by a special camera, secondly transmit the image obtained to the computer by image acquisition system, thirdly process and analyze the image in special particle size analysis software, finally output the result through monitor and printer. It is characterized by direct-image, direct-view, accuracy, broader range, automatic recognition, statistics and quantification. In addition to examining the shape of the particle, it gives particle size distribution and the ratio of length to diameter (L/D), etc. therefore it becomes a new method for particle size test in scientific research and production field.

Technical Specification

  1. Size Range:1-3000μm
  2. Max magnifying power: 4000X(optical + electric amplification)
  3. Max Resolution:0.1μm
  4. Repeatability: <3% (CRM D50)
  5. Particle size distribution type: volume, area, number, L/D.
  6. Output Items: Results output: Original parameter (including sample information, test information, etc.), analysis data (including differential distribution, cumulative distribution, average diameter, the ratio of length to diameter (L/D), etc.), image (differential particle size distribution histogram and cumulative distribution curve etc.)

Principle

  1. Calibration Method: Calibrate the size of each pixel by standard special graduated scale of the microscope, after that measure the size of the image by calculating the number of pixels. The unit is um.
  2. Test principle: Through the statistics of the number of the particle and the number of the pixels occupied by each particle, the area of equivalent circle and the volume of equivalent sphere are obtained, therefore diameter of equivalent circle area, diameter of equivalent sphere volume and L/D, etc. are obtained.
  3. How does GA-1600 measure size of particles: It uses 100 pixels as one unit for convenient calculation. Table 1 shows particle diameter of 100 pixels under different objective amplification
No.Objective Ampl.Pixel Num.Particle Dia.(um)Remarks
14100250

Corresponding particle Dia. is according to calibrated value.

210100100
32010050
44010025
56010016.7
610010010

4. System configuration(as per quotation)

No.DescriptionModelSpecificationQty.Remarks
1MicroscopeNikon YS100Max Ampl.16001

Necessary

2Data CCDHV2001UC192M pixels1
3SoftwareBT-1600Data acquisition and process1
4ComputerWindows system1

Optional

5Printerany model1
6UltrasonicSK-1200H50W1

How to process

  1. Image Process

2. Image analysis

BT-1600 Image Particle Size Analyzer suppliers    BT-1600 Image Particle Size Analyzer manufacturers