Automatic cleveland open cup flash point Tester
-Print data automatically
-Having prompt function to modify operation mistake
Port: Shanghai China
Minimum Order Quantity: 1 Set/Sets
Supply Ability: 1000 Set/Sets per Month
Payment Terms: L/C,T/T
Contact SupplierPackaging Detail: Carton
Delivery Detail: 15 days after received TT
Automatic cleveland open cup flash point tester
Model:LH-904
Description
LH-303 Automatic Cleveland Open Cup Flash Point Tester is designed and made as per GB/T3536 “Test Methods for Flash Point and Fire Point of Petroleum Products (Cleveland Open Cup Methods)”. It is suitable to determine open cup point of petroleum products, excepting fuel oils and petroleum products, whose flash point are lower than 79 °c. It is widely used in the fields of railway, aviation, power plant, petroleum industry, and scientific and research department.
Referring to ASTM D92.
Feature
1. Simulate and track to indicate function curve of temperature against test time.
2. Having prompt function to modify operation mistake.
3. Having test date and time.
4. Can calibrate barometric pressure automatically and calculate modification value.
5. Differential detection, can calibrate system deviation automatically.
6. Can open cover, ignite, detect, and print data automatically.
7. Test arm can be raised and lowered automatically.
8. Ignited by electron; force wind cooling.
Technical parameters
1. Power supply: AC 220 V±10%, 50 Hz;
2. Temperature measurement:
(1) Full scale: Ambient temperature to 300 °C;
(2) Repeatability: ≤8°C;
(3) Reproducibility: ≤16°C;
(4) Resolution: 0.1°C;
(5) Precision: 0.5%;
3. Basic parameters:
(1) Temperature rising speed: as per GB/T3536;
(2) Ignition mode: by an electric flame (gas flame is 3 to 4 mm);
4. Ambient temperature: 10~40°C;
5. Relative humidity: ≤80%;
6. Total power consumption: not more than 500W.