1.Atomic Absorption Spectrometer.2.optical equipment,refractometer digital ph meter,laboratory instrument,medical equipment,lab
Port: Beijing,Shanghai,Guangdong
Minimum Order Quantity: 1 Piece/Pieces
Supply Ability: 100 Piece/Pieces per Week
Payment Terms: L/C,D/A,D/P,T/T,Western Union,MoneyGram
Contact SupplierPackaging Detail: Export Standard Packing
Delivery Detail: Normally within 4-7days can shipment goods.
A & E Lab.
Main Products: Atomic Absorption Spectrometer (AAS), Fluorescence Spectrophotometer, Bio-Spectrophotometer, UV Vis Spectrophotometer, Visible Spectrophotometer, Flame Photometer, Whiteness Meter, Turbidity Meter, Colorimeter, Gloss Meter, Tintometer, Automatic Polarimeter, Auto Melting Point Apparatus, Automatic Refractometer, Abbe Refractometer, PH/DDS/TDS Meter, etc.
Features:
Fully Automated
² The flame and graphite furnace is integrated into one instruments.The change over from one technique to another is carried out by simple key strokes within the software.
² The instruments have a motorised 8 hollow cathode lamp turret which allows the automatic positioning and optimization of each hollow cathode lamp by the software.
² The spectral bandwidth is automated and is available with choice of five slit sizes.Advanced Graphite Furnace
² Horizontally heated platform graphite tube.The unique design of graphite furnace reduces the chemical interference effects and memory effects
by uniformly heating the graphite electrode.
² The computer controlled heating program all.Iows the users to select the best heating program for analysis.
² The optical temperature during the atomization stage ensures the rapid heating and rapid analysis.This helps
to extend the life of graphite tube and enhance the analytical accuracy.
Technical Specifications
Model | AE-AA990 |
Optic System | |
Wavelength Range | 190nm-900nm |
Monochromator | Czemy-Turner Configuration |
Spectral Bandwidth | 0.1nm,0.2nm,0.4nm,1.0nm,2.0nm(5steps. with automatic change over) |
Wavelength Accuracy | ±0.25nm |
Wavelength Repeatability | 0.15nm |
Baseline Stability | 0.005A/30min |