Home  >  Optical Instruments  >  Spectrometers

Advanced Spectrometer(AE-UV1801) - AE-UV1801

1.Advanced Spectrometer:190-1100nm,bandwith 0.5,1,2,4nm., 2. Double Beam,Flame,Optical,Spectrum Spectrometer., 3. Labs - more info check here: http://www.buymeasuringtools.com/Advanced-Spectrometer-AE-UV1801-10302865/

Quick Brief

Place of Origin: China (Mainland) Brand Name: A & E Lab Model Number: AE-UV1801 Wavelength Range: 190-1100nm Measurement Range: 0.5nm,1nm,2nm,4nm

Product Features:

1.Advanced Spectrometer:190-1100nm,bandwith 0.5,1,2,4nm.

2. Double Beam,Flame,Optical,Spectrum Spectrometer.

3. Labs

Business Terms:

Minimum Order Quantity: 1 Piece/Pieces

Supply Ability: 100 Piece/Pieces per Week

Payment Terms: L/C,D/A,D/P,T/T,Western Union,MoneyGram

Contact Supplier
Advanced Spectrometer(AE-UV1801)
Share to: Facebook Twitter LinkedIn Google+

Product Packagings & Delivery Terms:

Packaging Detail: Export Standard Packing

Delivery Detail: Normally within 7-10days can shipment goods.

Specifications and Product Details:

A & E Lab.

Main Products: Atomic Absorption Spectrometer (AAS), Fluorescence Spectrophotometer, Bio-Spectrophotometer, UV Vis Spectrophotometer, Visible Spectrophotometer, Flame Photometer, Whiteness Meter, Turbidity Meter, Colorimeter, Gloss Meter, Tintometer, Automatic Polarimeter, Auto Melting Point Apparatus, Automatic Refractometer, Abbe Refractometer, PH/DDS/TDS Meter, etc.

Features:

# With embedded computer and optimized optical system,convenient and easy to use.

# 320×240 dots array large screen graphic LCD display with abundant instructions,and curve and table directly

displayed on screen.

# Basic photometric mode, Quantitative analysis Qualitative analysis, Kinetics test, DNA/Protein Test and

Multi-wavelength are available.

# Alternative fitting and correction methods for Quantitative Analysis.

# Re-scaling axes, curve smoothing,combination,zooming and overlap, 1st to 4st derivative for Qualitative analysis.

# Up to 1000 test data can be saved in local memory.

# Data output by optional built-in thermal printer or USB2.0 printer driver and PC software available.

Technical Specifications:

ModelAE-UV1801AE-UV1801S
Wavelength Range

190 - 1100nm

Spectral Bandwidth2nm or 1nm0.5nm,1nm,2nm,4nm,5nm Variable
Optical System

Double Beam & Czerny-Turner; Grating 1200 line/mm

Wavelength Accuracy

+/-0.1 nm(656.1 nm D2), +/-0.3 nm(Whole wavelength range);

Wavelength Repeatability

+/-0.2nm

Wavelength Resolution

0.1nm

Photometric Range

­0.3 to 3A ; 0-200%T; -9999 to 9999C;

Photometric Accuracy

+/-0.002A, 0.5A , +/-0.3%(0~100%T)

Photometric Repeatability

+/- 0.001A, 0.5A , +/- 0.15%T(0~100%T)

Stray Light

less than 0.05%T (220 nm, NaI; 340 nm, NaNO2)

Baseline Flattness

+/- 0.001A

Stability

+/- 0.0005A/hr, 500nm, 0A

Photometric noise

+/- 0.0003A

Data Output

320×240 Dots Array Graphic LCD;

USB2.0 printer driver(Optional);

USB1.1 Connecting to computer;

Advanced Spectrometer(AE-UV1801) suppliers    Advanced Spectrometer(AE-UV1801) manufacturers